Thư viện tri thức trực tuyến
Kho tài liệu với 50,000+ tài liệu học thuật
© 2023 Siêu thị PDF - Kho tài liệu học thuật hàng đầu Việt Nam

HPLC A Praactical User''''S Guide Part 4 pdf
Nội dung xem thử
Mô tả chi tiết
times 2-fold while increasing efficiency only by 1.4-fold due to increased
diffusion.
Finally, we have variables affecting efficiency that can be controlled at the
time of the run. These are pump flow rate, extracolumn volumes in the instrument used, and the method of calculation. Flow rate is the major efficiency
variable that I use during methods development. Generally, halving the flow
rate will increase separation around 40%. I do much of my scouting at
2.0 mL/min, knowing that I can improve separation by dropping to 1.0 mL/min.
Plotting of efficiency versus flow rate shows that each diameter of packing has
its own optimum flow rate. Efficiency decreases at higher flow rates. In the
microparticulate packings, large packing diameters show a more rapid loss of
efficiency with increasing flow rate than do smaller packings.
Decreasing extracolumn volumes is critical to HPLC success. The most
important volumes are those immediately adjacent to the column: zero-deadvolume end-fittings, inlet and outlet tubing diameters, and detector cell
volumes. From the time the sample enters the injector until it exits the detector, nothing must add increased mixing space. Tubing from injector to column
must be 0.010 in for 5-mm and 10-mm packings with tubing lengths no more
than 4–6 in for the 5-mm. Use 0.007-in tubing about 3 in long or less for 3-mm
packing. Zero-dead-volume endcaps and connectors must be prepared correctly, so that tubing ends butt firmly against the fitting. We covered the preparation of compression fittings in Chapter 3, but if you find efficiency drops after
you change a fitting, check the dead-volume fit. For detector cells, the rule of
thumb is 8–12mL; anything larger acts increasingly as a mixer for your already
separated bands.
Tubing volumes outside the critical injector-detector range are important
only if you are doing recycling or collecting samples. Pump-to-injector tubing
is generally 0.020-in; vents, flush valve, etc. may use 0.04-in. Be sure you know
what these look like and do not confuse them with injector tubing. In telling
tubing apart, 0.02-in and 0.01-in are the most difficult to tell apart. If you have
to look twice to make sure there really is a hole, it is probably 0.01-in. If you
are in doubt, put them next to each other. By comparison, 0.04-in tubing looks
like a sewer pipe.
There are many methods used to calculate efficiency. All methods give the
same results with ideal, Gaussian peaks. Real chromatography peaks tend
to tail on the backside of the peak (away from the injection mark on the
PARTITION 51
Table 4.1 Relationship of efficiency to flow rate
Efficiency Changes with Particle Size
Packing diameter (mm) Plates/meter Flow rate (mL/min)
10 30,000 1.0
5 50,000 1.5
3 100,000 2.5
chromatogram). When column problems occur they often tend to show up as
increased tailing. Calculation methods that use a peak width high on the peak
miss these changes and give artificially high efficiencies. The 5s method
described above is excellent for detecting early appearance of tailing. If you’re
planning on using a calculation using half-peak width, make sure there is some
method of measuring and correcting for peak asymmetry.
The retention factor, k′, also called the capacity factor, is the usual starting
point for methods development. The retention factor, as its name implies, is
basically a measure of how long each compound stays on the column; VO used
to determine k′ is usually only roughly measured; k′ is a simply a multiple of
the VO distance (see Fig. 4.5).
The major usable variable controlling k′ is solvent polarity. While temperature and column polarity also effect retention times, they do not show the
same direct, linear relationship for all peaks and are usually classed under the
separation factor (a).
Increasing the polarity difference between the stationary and mobile phases
increases the retention of compounds with polarities most like the column.
Compounds stick tighter and peaks will broaden through diffusion. Decreasing the polarity difference will make things come off faster and shoved
together. Peaks will be less resolved and sharper.
For example, for a polar silica column equilibrated with a mobile phase of
methylene chloride in hexane (nonpolar), you would dilute with more hexane
to increase the k′ of relatively polar components. Adding methylene chloride,
the more polar of the two solvents, would decrease k′s causing all components
to wash off faster.With k′ changes, peak position changes are proportional and
in the same direction. The order of resolved peaks will remain the same; unresolved peaks should begin to pull apart.
If in our model system, we had used 80% methylene chloride/hexane and
the red peak had partially overlapped the backside of the blue peak, we would
attempt to resolve it by reequilibrating in 40% methylene chloride/hexane and
reinjecting.We could expect that we should see two well-resolved peaks; if not,
we could go to a 20% mixture. More than likely, we would have overshot on
the first change and would have to fine-tune back toward the 80% mixture.
Simply by modifying the solvent polarity, we are able to increase or decrease
k′ and contract or spread our separation. This k′ development is our usual
starting point in methods development.
So far, I have referred only to “normal-phase” separations on polar
columns. However, around 80% of the separations in the literature are made
on “reversed-phase” columns. To understand these terms, we need a little
history.
52 SEPARATION MODELS
Figure 4.5 Retention factor equation.