II.1.7 Toluene, benzene, xylene
and styrene
by Xiao-Pen Lee and Keizo Sato
Introduction
Toluene, benzene, xylene and styrene are being widely used for chemical product materials,
solvents and constituents of adhesives and paints, and thus sometimes cause poisoning incidents by inhaling their gas at chemical product-manufacturing factories. Th e abuse of thinner
solvents, containing toluene, benzene and xylene, is now a social problem especially for young
people. Th ere are many crimes and poisoning incidents involving the thinner solvent abuse.
Recently, styrene leaking from new building materials is an object of interest, because it causes
the sick house syndrome.
Th e above four solvents are rapidly absorbed into human body by inhalation, per os
and percutaneously to cause various disturbances, such as suppression of the central nervous
system. In addition, benzene and styrene are being regarded as carcinogenic compounds.
For analysis of toluene, benzene, xylene and styrene, the conventional GC methods with
packed columns are being used [1–10]. In this chapter, a method of their analysis for human
blood using headspace wide-bore capillary GC is presented.
Reagents and their preparation
• Standard solutions: toluene, benzene or each isomer a
of xylene of special grade is dissolved
in methanol to prepare each 1 mg/mL solution.
• Internal standard (IS) solution: aniline of special grade is dissolved in methanol to give
10 mg/mL solution.
GC conditions
Column: an Rtx-BAC2 wide-bore capillary column (30 m × 0.53 mm i. d., fi lm thickness
2.0 µm, Restek, Bellefonte, PA, USA).
GC conditions; instrument: a Shimazu GC-14B gas chromatograph (Shimadzu Corp.,
Kyoto, Japan); detector: FID; column (oven) temperature: 30 °C (1 min) → 15 °C/min → 240 °C
injection and detector temperature: 240 °C; carrier gas: He; its fl ow rate: 5.0 mL/min.