87
Analytical
parameters
Limit of detection and limit
of quantification
Selectivity
The most important parameters for food analysis are:
• limit of detection (LOD) and limit of quantification
(LOQ)
• linearity
• selectivity
• qualitative information
The LOD and LOQ of an analytical system depend on the
noise and drift of the detection equipment. Absolute detector LOD can be determined by injecting a sample directly
into the detector. It is often expressed as minimum detectable level, which is sometimes defined as equal to the noise
level. However, the LOD depends not only on the detector
but may also depend on the oxygen content of the mobile
phase, the injection system, peak broadening on the column, and temperature differences among system components. Taking these factors into account, the LOD is defined
as 2 to 3 times the noise level. The LOQ is defined as 10 to
20 times the noise level. A UV detection system can be used
to measure quantitative amounts down to 500 pg per injection. The LOD can be as low as 100 pg for food compounds
such as antioxidants if detection wavelengths have been
optimized to match the extinction coefficients of as many
compounds as possible. Fluorescence and electrochemical
detectors operate in the very low picogram range. The LOD
of a mass spectrometer connected to HPLC equipment
depends on the type of interface used. Instruments with
electrospray interfaces can detect down to the picogram
range. Refractive index detectors normally are appropriate
above 500 ng.
We define the selectivity of a detection system as the ability
to select only those compounds of interest in a complex
matrix using specific compound properties. A detector is
selective if it does not respond to coeluting compounds that